Operation For Coefficient Of Friction Tester

Standards :

ASTM D1894

ISO 8295

TAPPI T816

GB 10006

  • OPERATION
    1. Take 2 samples, move the test sample (a square of 63.5mm on one side) and paste it on the metal slider with double-sided tape, and fix the test sample (a rectangle of 130mm wide and 250mm long) to the fixed table with double-sided tape.
    2. Place the sliding block with fixed moving pattern in the center of the fixed sample, and make the test direction of the two samples parallel to the direction of the sliding block, and keep the force measuring system without force.
    3. Set parameters and click “Start” button to start the experiment.
    4. When the number of tests reaches the set number, the test group will automatically end.
    5. After the experiment, turn off the power.
  • Note: After the device is finished, turn off the power and cover with dust cloth. The dust cloth should be clean and tidy, and free of dust.
  • Watch the video for details
Scroll to Top